HI98308 เครื่องวัดค่า Conductivity ในน้ำกลั่น DI และ น้ำ RO มีความแม่นยำขนาดพกพาที่จะให้คุณตรวจสอบค่าคอนดักติวิตี้ในน้ำได้รวดเร็วและถูกต้อง ใช้ในการตรวจสอบความบริสุทธิ์ของน้ำ น้ำปราศจากแร่ธาตุในห้องปฏิบัติการหรืองานด้านสภาพแวดล้อมอุตสาหกรรม คุณภาพสูงและคู่มือการใช้งานภาษาไทย
– ช่วงการวัดสูงถึง 99.99 uS/cm
– ใช้งานง่ายมาก
– รวดเร็วและแม่นยำ
– การอ่านค่าที่ชดเชยอุณหภูมิ
– ราคาไม่แพง
The HI98308 features an amperometric graphite electrode that provides better repeatability in measurements since they do not oxidize. An amperometric measurement of EC is based on Ohm’s Law, I = V/R, where R depends on the distance between two pins and their surface. Oxidation changes both the distance and surface, which will affect accuracy. The non-oxidizing graphite pins are able to provide an optimal surface for accurate, dependable results. When calibration is needed, simply submerge the electrode tip into calibration solution and adjust the trimmer on the side of the tester.
The HI98308 uses a conductometric measurement to determine the purity of water. The HI98308 Pure Water Test (PWT) enables users to check the purity of distilled or demineralized water in laboratory or industrial environments. PWT is suited for fields such as printed circuit board washing, laundry, steam cleaning, checking car battery water and all areas where distilled, demineralized or pure water is used. This tester is housed in durable casing that provides excellent protection against harsh industrial environments.
Electrolytic conductivity, abbreviated as EC, is a measurement made in which electrical charges on atomic or larger sized particles in a medium are moved under the influence of a potential difference. EC is a measure of concentration, however, it is non-specific for ion type. An ion is a charged particle present in the solution that contributes to the current flow. Ions are formed when a salt such as sodium chloride is dissolved in water to form particles having electrical charges. Sodium chloride, separates into Na+ and Cl-. This is a simplified definition as the measurement is affected by many things such as the type of ionic compound(s) dissolved in the water; the ions mobility, the solution viscosity, temperature as well as concentration.
PWT – Conductivity tester for distilled or ultrapure water (up to 99.9 uS / cm).
In the laboratory distilled water for many studies is needed daily. The handy PWT easily and quickly checks the quality of pure water. Low maintenance required. PWT has a probe of non-oxidizable graphite, and a built-in temperature sensor for temperature compensated readings.
- Measuring range up to 99.99 uS / cm
- Very easy to use
- Fast and accurate
- Temperature-compensated readings
- Inexpensive
Manual : Download
Specification :
SKU | HI98308 |
Product Name | Pure Water Tester (0.0 – 99.9 µS/cm) – HI98308 |
Quote Required | Yes |
EC Range | 0.0 to 99.9 µS/cm |
EC Resolution | 0.1 µS/cm |
EC Accuracy | ±2% F.S. |
EC Calibration | manual, one point |
EC/TDS Temperature Compensation |
automatic from 0 to 50°C (32 to 122°F) with β=2%/°C typical |
Battery Type/Life | 1.5V (4) / approximately 250 hours of continuous use |
Environment | 0 to 50°C (32 to 122°F); RH max 95% non-condensing |
Dimensions | 175 x 41 x 23 mm (6.9 x 1.6 x 0.9’’) |
Weight | 95 g (3.4 oz.) |
Ordering Information | HI98308 is supplied with protective cap, calibration screwdriver (HI98308), batteries and instructions. |